Improvement of the Shock Response Level of Shaker by Changing Linking Characteristic
Received:November 08, 2015  Revised:April 15, 2016
View Full Text  View/Add Comment  Download reader
DOI:10.7643/issn.1672-9242.2016.02.006
KeyWord:fixture  linking characteristic  shaker  shock response spectrum
        
AuthorInstitution
SHEN Zhi-qiang Beijing Institute of Spacecraft Environment Engineering, Beijing , China
ZHU Zi-hong Beijing Institute of Spacecraft Environment Engineering, Beijing , China
GUO Jian-long Beijing Institute of Spacecraft Environment Engineering, Beijing , China
Hits:
Download times:
Abstract:
      Objective To improve the shock response level of shaker. Methods The reasons that influenced the shock response level of shaker were analyzed. Then the method to improve the shock response level of shaker by changing linking characteristic was put forward, while the theoretical analysis and the test demonstration were carried out. Results The SRS level with 160 kN shaker could be improved from 1000g to 2000g by using the table with harmonic response. The SRS level under load of 1600g could be conducted. The shock response was improved. Conclusion The shock response level of shaker can be improved by changing linking characteristic, at the same time, the fixture of vibration should be avoided. The effect of SRS can be better with the fixture of vibration which are designed as required.